Anesthesia Ventilator Machine MSLGA13

Original price was: $16,990.00.Current price is: $15,990.00.

SKU:MSLGA13

Specifications

1 Operating mode: circulatory closed, semi-closed and semi-open.
2 Gas requirement: medical oxygen and laughing gas with a pressure ranging from 0.3MPa to 0.5 MPa.
3 The maximum error of the pressure testing device should not exceed ± (4% of the full scale reading + 8% of the
actual reading).
4 For oxygen and laughing gas, a special pressure regulator with a safety valve will be provided. The exhaust pressure
of the safety valve should not be higher than 6 kPa.
5 Indication range of the oxygen and laughing gas flow meter: 0.1 L/min ~ 10 L/min.
When the flow rate ranges from 10% of the full scale value to 100%, the scale accuracy should be within ±10%
of the indicated value.
6. The flow meter is equipped with an oxygen-laughing gas proportional control device.
When the oxygen concentration in the mixed gas N2O/O2 conveyed by the anesthesia
machine is no lower than 20% (V/V) or the FiO2 is lower than 20%, the machine will emit an
alarm.
7. When the oxygen pressure of the anesthesia machine is 0.20MPa±0.05MPa, the machine raises a low gas source
pressure alarm which is a high priority alarm, and shuts off the laughing gas conveyed to the common gas outlet.
8. Oxygen Flush:25~75 L/Min;
9. Anesthetic gas concentration adjustment range of vaporizer: 0~5%, relative error ±20 %.
10. The exhaust pressure of the safety valve of the respiratory circuit is no higher than 6 kPa.
11. Anesthesia ventilator
11.1 Respiration Mode:IPPV,SIPPV,Manu
11.2 Respiration Frequency:4~40bpm
11.3 I/E Ratio:1:1.5~1:4
11.4 Tidal Volume:50~1500mL
11.5 Ptr:-10~10hPa
11.6 Controlled ventilation and assisted ventilation switching time: 6s
11.7 Maximum safety pressure:  ≤ 12.5 kPa.
11.8 Pressure limit range: 1~6 kPa
11.9 Airway pressure alarm: adjustment range of upper alarm limit: 0.3kPa ~ 6 kPa, permissible error ±0.2 kPa, or ±15%
(whichever is greater), the machine should immediately raise a high level alarm when the airway pressure rises to the alarm
value; the lower alarm limit ranges from 0.2 to 5 kPa and the permissible error is ±0.2 kPa or ±15% (whichever is greater).
The machine should immediately raise a medium level alarm when the airway pressure drops to the alarm value and
such condition lasts 4 to 15s.

Warranty
All products listed on our website have 2 year warranty. Lifetime support, technical service available all around the world.

Expected Delivery Times
United States & Canada: 5-7 Days by FedEx
Europe: 3-5 Days by DHL or UPS
Asia & Africa: 7 Days by DHL or UPS
Middle East: 9 Days by DHL or UPS
Austuralia: 5 Days by DHL or TNT
Central & South America: 7-9 Days by DHL or UPS

Product Origin
Our company has warehouses in China, Turkey, Korea, USA. Main operating center located in London, UK.

Technical Service
If you have a problem with the tool you have purchased, follow the link and complete the form immediately. After we receive your request, our team will get in touch with you within 24 hours for maintaince.

Configuration of Power Tools
Standard configuration of battery powered power tools: 2 batteries, charger, briefcase, user manual, warranty card, aseptic kit. Chuck and key for drill groups, Blades for saw groups included within products.

Reviews

There are no reviews yet.

Write a review

Your email address will not be published. Required fields are marked *

Back to Top